9 Horizontal Arm Brake Cable Removal - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
3 9
Horizontal Arm
Brake Cable Removal
ILLUSTRATION 3-7
BRAKE AND CABLE ASSEMBLY
PULLEY CABLE
PULLEY SHAFT
BEARING
PULLEY
WASHER
PULLEY BAR
MOUNTING BLOCK
PULLEY CABLE
10–32X5/8 INCH SCREW
10–32X1 INCH SCREW
1. Remove horizontal arm brake and cable assembly. Refer to Section 3-7.
2. Remove cable clip from bar support assembly by removing one machine screw.
See Illustration 3-7.
MASTER LINK
CABLE BAR
PULLEY BAR
PULLEY CABLE
CABLE BAR
BRACKET
6–32X5/8 INCH SCREW
3. Remove anchor nut from brake cable by turning counterclockwise. This nut has
been secured with set screws and Loctite.
4. Remove cable anchor end from bar support.
5. Slide ball end of cable from slot in cable bracket and remove from path around
pulley end on mounting plate.
6. Remove the brake cable from brake and cable assembly.
(MODEL 2169360, 2236420 & 2275938 SERIES)
CABLE BAR
BRAKE CABLE
MOUNTING BLOCK
CABLE ANCHOR
6–40 1/8 INCH SCREW
10–32X1/2 INCH SCREW
BEARING
SPROCKET
BEARING
CHAIN SHAFT
BRAKE
3-12
AMX-4+ SERVICE
DIRECTION 2173225-100
BRAKE CHAIN
MASTER LINK
BRAKE WIRING