10 3 3 Collimator Cable - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5

10 3 3 Collimator Cable

ILLUSTRATION 10-9
COLLIMATOR CABLE
ROUTE TO FIRST STRAP
ANODE
CABLE
ILLUSTRATION 10-10
CATHODE AND COLLIMATOR CABLES
1. The collimator cable starts at the collimator, passes over the anode cable and
goes to strap point J" where it is joined to the cathode cable. There is a large loop
of cable before reaching strap point J" as shown in Illustration 10-9.
TUBE
CATHODE
2. Continue routing and ty rapping the collimator cable along the cathode cable to
strap point C." See Illustration 10-10.
CATHODE
CABLE
10-8
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
ANODE
COLLIMATOR CABLE
COLLIMATOR
CABLE
AMX-4+ SERVICE
CATHODE
CABLE
STRAP POINT  J"