GE MEDICAL SYSTEMS
REV 5
7. Return arm to upright position.
8. Slide the center and inner tubes from the outer tube.
9. Place center and inner tubes in upside down position.
10. Remove four screws holding latch block to center tube. Remove block. This pro
vides access to the mounting plate holding screw. See Illustration 3-6.
11. Remove mounting plate capscrew holding brake and cable assembly to the center
tube.
12. Remove the two brake capscrews holding the brake & cable assembly to the rear
of the center tube.
13. Remove two or three screws (depends on unit you have) holding the connector
bracket and ground wire to the center tube.
14. Remove two retaining rings from inner shaft on center tube. Slide shaft out per
mitting removal of two spacers and two bearings.
15. Carefully slide the brake and cable assembly, and the inner tube from the center
tube.
16. Remove two capscrews holding the brake and cable assembly bracket to the inner
tube.
17. Remove brake and cable assembly from inner tube.
3-10
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE