Servicing The High Voltage Transformer; High Voltage Transformer Removal - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
SECTION 9

SERVICING THE HIGH VOLTAGE TRANSFORMER

9 1
High Voltage
Transformer Removal
ILLUSTRATION 9-1
HIGH VOLTAGE TRANSFORMER
BINDING
HEAD SCREW
HEX WASHER
HEAD SCREW
This section provides procedures covering the removal and installation of the high
voltage transformer.
1. Shut power off by turning the circuit breaker OFF.
2. Remove front cover by carefully pulling cover from four ball stud retainer sock
ets.
3. Disconnect anode and cathode cables from high voltage transformer.
4. Remove two cap screws holding rear cover on base assembly and remove cover.
5. Remove cassette drawer. Refer to Section 5-1.
6. Remove cassette drawer magnetic strip lock by removing three binding head
screws and flat washers.
7. Remove high voltage transformer terminal cover by removing two binding head
screws and washers. See Illustration 9-1.
HV TRANSFORMER
DRIVE ELECTRONICS MODULE
9-1
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
BASE ASSEMBLY
MAINFRAME ASSEMBLY
AMX-4+ SERVICE