Lower Brake Removal - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
4 7

Lower Brake Removal

ILLUSTRATION 4-12
LOWER BRAKE AND BEARINGS
1. Remove vertical column. Refer to Section 4-1.
2. Remove retaining ring from bottom of column base shaft. See Illustration 4-12.
UPPER COLUMN BEARING
LOWER COLUMN BEARING
3. Pull lower column bearing from column base shaft.
4. Remove screws and washers securing lower brake to column bearing plate.
5. Remove lower brake from column base shaft.
4-21
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
COLUMN BASE SHAFT
COLUMN BEARING PLATE
AMX-4+ SERVICE