4 Light Field Adjustment - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
1 4
Light Field Adjustment
1 4 1
Adjustment Method 1
ILLUSTRATION 1-5
APPEARANCE OF X-RAY
FIELD AND LIGHT FIELD
ALIGNMENT ON EXPOSED
FILM USING TEST PATTERN
The x-ray field and light field must be properly aligned to meet HHS requirements.
Two methods for accomplishing this are given below.
Method 1, Section 1-4-1, is only used if the Light Field to X-Ray Field Test Pat
tern," 46-198466P1, from an HHS Field Test Kit, 46-177372G1, is available.
Method 2, Section 1-4-2, uses lead strips to indicate the light field areas.
1. Place loaded cassette (14  17 inches suggested) on floor directly beneath colli
mator.
2. Adjust the collimator so that it is parallel to the cassette.
3. Set SID to 40 inches.
4. Place light field to x-ray field test pattern on top of cassette.
5. Turn on field lamp.
6. Adjust collimator blades so light field exactly covers the 8  10 inch (20.32 x 25.4
cm) rectangle on the test pattern. See Illustration 1-5.
A
1-6
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
C
B
D
AMX-4+ SERVICE
FILM
TEST PATTERN
8 X 10 INCH TEST
RECTANGLE
(LIGHT FIELD)
X-RAY FIELD
HORIZONTAL
DIFFERENCE
SECOND
EXPOSURE
VERTICAL
DIFFERENCE

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2169360 series2236420 series2275938 series

Table of Contents