14 Column Support Assembly Removal; 15 Column Support Assembly Installation - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
4 14
Column Support
Assembly Removal
ILLUSTRATION 4-15
COLUMN SUPPORT ASSEMBLY
4 15
Column Support
Assembly Installation
1. Remove vertical column assembly. Refer to Section 4-1.
2. From beneath the bare assembly, remove capscrews and washers securing col
umn support to bottom of base assembly. See Illustration 4-15.
3. Remove hex head capscrews, lockwashers and flat washers securing the column
support to the base assembly.
4. Remove column support from base assembly.
1. Place column support in base assembly. Orient opening in side of column support
so it is opposite terminal board in base assembly. Line up mounting holes. See
Illustration 4-15.
2. Secure the column support to the base assembly with hex head capscrews, lock
washers and flat washers. Torque to 100-120 pound-inches (11.2 to 13.4 N-m).
3. Secure bottom of column support to bottom of front tongue in base assembly with
cap screws and washers. Torque to 100-120 pound-inches (11.2 to 13.4 N-m).
4. Install vertical column assembly. Refer to Section 4-2.
4-26
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
CAPSCREWS(8)
LOCK WASHERS(8)
FLAT WASHERS(8)
TORQUE TO 100-120 LB. INS.
(11.2 TO 13.4 N-M)
COLUMN SUPPORT
BASE ASSEMBLY
LOCKWASHERS (4)
CAPSCREWS(4)
TORQUE TO 100-120 LB. INS.
(11.2 TO 13.4 N-M)
AMX-4+ SERVICE