Yoke Mounting Plate Assembly Removal - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
3 3
Yoke Mounting Plate
Assembly Removal
Right Mounting Plate Assembly
ILLUSTRATION 3-2
RIGHT MOUNTING PLATE ASSEMBLY
1/4-20 X .5 HEX SOCKET FLAT
HEAD CAPSCREW
10-32 X .500 HEX SOCKET
FLAT HEAD CAPSCREW
MOUNTING PLATE
1. Remove collimator. Refer to Section 1.
2. Remove the x-ray tube. Refer to Section 2.
3. Remove decal from cover on end of right tube support to access the cover mount
ing screws.
4. Remove four screws holding cover to mounting plate shaft.
5. Remove cover, two spring washers, two thrust washers and thrust bearing from
mounting plate shaft.
6. Slide shaft, disk, and mounting plate from right tube support.
7. Remove two small bearings and their shaft from right tube support.
8. Remove four screws holding shaft and disk to mounting plate.
9. Remove shaft and disk from mounting plate.
DISK
MOUNTING PLATE SHAFT
BEARING
(MODEL 2169360, 2236420 & 2275938 SERIES)
1/4-20 X .625 HEX SOCKET FLAT
HEAD CAPSCREW
SHAFT
THRUST WASHER
COVER
10-32 X .500 HEX SOCKET
FLAT HEAD CAPSCREW
3-3
AMX-4+ SERVICE
DIRECTION 2173225-100
RIGHT TUBE SUPPORT
THRUST BEARING
SPRING WASHER
DECAL