10 1 2 Cord Reel Removal (IEC Models)

ILLUSTRATION 10-2
CORD REEL REMOVAL/INSTALLATION
1. Shut power off by turning circuit breaker OFF.
2. Open top cover. Refer to Section 5-5.
3. Disconnect cord reel leads at main terminal block. See Illustration 10-2.
4. Remove two internal hex cap screws retaining grommet in cut-out in mainframe
and remove cable and grommet.
5. Remove two hex washer head screws and lock nuts holding cord reel to side of
low voltage power supply housing.
6. Loosen screw in strain relief bracket located behind cord reel.
7. Lift off cord reel pulling leads through strain relief bracket.
BALL
GROMMET
HEX WASHER
HEAD SCREW
CORD REEL
10-2
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
STRAIN
RELIEF
BRACKET
AMX-4+ SERVICE
LOW
VOLTAGE
POWER
SUPPLY
HOUSING
MAINFRAME
MAIN
TERMINAL
BLOCK