GE MEDICAL SYSTEMS
REV 5
3 6
Horizontal Arm
Assembly Installation
Note:
CAUTION
The counterweight must be at the bottom of the column or otherwise immobi
lized.
1. Mount horizontal arm assembly on carriage assembly flanges. See Illustration
3-5.
2. While supporting the arm, secure it firmly in place with six mounting screws.
Torque to 110 pound-inches (12.3 N-m).
3. Route brake cable along the anode cable to strap point on side of the column.
Secure brake cable to the anode cable with ty-raps.
4. Drape the cable from the side of the column to the terminal box on the back of
the column. Insert the brake cable with a Heyco bushing up through the bottom
of the terminal box.
5. Connect brake cable leads to terminal block in accordance with connection data
recorded and tagged on leads.
6. Position back cover on terminal box and secure with two screws.
Equipment Damage!
The horizontal arm must be secured before installing components on the
arm. Do not rely on latch alone to hold it in place. With weight removed from
the arm, the counterweight can descend rapidly and the arm ascend rapidly
possibly causing damage to the unit and personal injury. Use Counterweight
Lift Tool 46-303491G1.
7. Firmly grasp the horizontal arm and using your weight to counterbalance the
counterweight, approximately 75 pounds (34 kg), carefully move it down to the
top of the unit and into parked position. Do not rely on latch alone to hold the
arm in place. Check that the arm is locked down before slowly removing your
weight and releasing it.
8. Re mount the x-ray tube and collimator assembly per Illustration 3-4.
9. Check that the arm moves freely up and down the column.
3-8
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE