Servicing The Batteries; 1 Remove Battery Cover - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
SECTION 6

SERVICING THE BATTERIES

WARNING
6 1
Remove Battery Cover
Note:
ILLUSTRATION 6-1
BATTERY COMPARTMENT ACCESS
PERSONAL INJURY !
REMOVE ALL JEWELRY AND RINGS WHEN HANDLING THE
BATTERIES OR REACHING IN THE BATTERY COMPARTMENT. USE
EXTREME CAUTION, HIGH AMPERAGE AND VOLTAGE IS PRESENT.
USE INSULATED TOOLS FOR WORKING ON BATTERIES. WEAR
SAFETY GLASSES.
During removal:
D
Reference to the right and left sides of battery compartment are relative to
the view shown in Illustration 4.
D
Cable numbers (#s) indicated in Illustration 6-3 and in this instruction are
marked on the parts.
D
Insulator numbers are shown for convenience in description, but are not
marked on the parts.
D
There are no Insulators #3 and #7.
D
Battery numbers may or may not be marked on batteries. Mark batteries
with numbers if desired.
D
Disconnect cables only in the order instructed, and only when instructed.
D
Don't remove paper backing from any insulating separator pad.
1. Place the unit on a level floor. Extend the horizontal arm and rotate it to the front
end (vertical column end) of the unit. See Illustration 6-1.
Distribution of weight is necessary to remove the battery compartment cover.
CIRCUIT
BREAKER
COVER CAPSCREWS
6-1
(MODEL 2169360, 2236420 & 2275938 SERIES)
BATTERY
GROUND STUD
CABLE MARKED #4
CABLE MARKED #2
BASE ASSEMBLY
BATTERY
COMPARTMENT
AMX-4+ SERVICE
DIRECTION 2173225-100