Install Drive Wheels
Note:
WARNING
Note:
DRIVE BELT
DRIVE
WHEEL
This procedure covers the installation of either the right or left drive wheel.
PERSONAL INJURY!
THE UNIT CAN BE UNSTABLE WHEN ON BLOCKS AND TIP OVER.
USE EXTREME CAUTION WHEN PERFORMING THIS OPERATION.
Apply light coat of grease (Lubriplate Fiske 630-A-A) to each end of axle.
1. The unit is on blocks approximately 10 inches (25 cm) above the floor.
2. Place the caliper assembly on the wheel axle.
3. Install the drive belt on the drive wheel pulley.
4. Lift the drive wheel through the bottom of the base assembly and into position
in the notch in the bottom of the wheel mounting support. See Illustration 7-7.
5. Attach the wheel and caliper assembly to the wheel mounting support with four
hex socket head capscrews. Do not tighten in place.
(MODEL 2169360, 2236420 & 2275938 SERIES)
HEX SOCKET
HEAD CAP SCREW
7-7
AMX-4+ SERVICE
DIRECTION 2173225-100
WASHER
RIGHT SIDE AXLE
DISC/CALIPER
ASSEMBLY