SERVICING THE CASSETTE DRAWER, COVERS, HORIZONTAL ARM LATCH AND HANDSWITCH

5 1
Cassette Drawer Removal
ILLUSTRATION 5-1
CASSETTE DRAWER
HANDLE
CASSETTE
DRAWER
SCREWS
5 2
Cassette Drawer Installation
This section provides procedures covering the removal and installation of trim covers.
1. Remove rear cover from unit by removing two screws securing it to base assem
bly. See Illustration 5-1.
2. Remove five binding head screws and flat washers securing cassette drawer hinge
to base assembly.
3. Pull cassette drawer to open position. It is retained by magnetic catch. Lift
drawer up and out of base assembly.
SCREWS (2)
REAR COVER
1. Place cassette drawer in position on base assembly.
2. Attach cassette drawer hinge to base with five binding head screws and flat wash
ers.
3. Install rear cover to base assembly with two screws.
4. Close cassette drawer. Magnetic catch will hold it in its place.
5-1
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE
BASE ASSEMBLY