Install Pads; Bumper Switch Removal; 1 4 Bumper Switch Installation - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
7 1 2

Install Pads

7 1 3

Bumper Switch Removal

ILLUSTRATION 7-2
BUMPER SWITCH
7 1 4
Bumper Switch Installation
1. Install the middle piece of the front bumper using 10-32 x 1.25 inch bolts and
flat washers. See Illustration 7-1.
2. Install each of the two corner pads with 10-32 x 1.25 inch bolts for each pad.
1. Remove pads. Refer to Section 7 1 1.
2. Remove actuator bracket to release tension on switch actuator. See Illustra
tion 7-2.
3. Remove two binding head screws holding switch on end of bumper track.
4. Remove switch from bumper assembly.
BUMPER
TRACK
SWITCH
BINDING
HEAD
SCREWS
ACTUATOR
BRACKET
WITH SPRING
1. Shut off power.
2. Install switch on end of bumper track and secure with binding head screws. See
Illustration 7-2.
3. Install actuator bracket to place tension on switch actuator.
4. Verify that bumper switch is activated by spring tension on switch actuator
bracket.