GE MEDICAL SYSTEMS
REV 5
5 3
Side Covers Removal
Note:
ILLUSTRATION 5-2
SIDE COVERS
RIGHT COVER
6-32 X 1/2 BLACK
OXIDE SCREW
BALL STUDS
FRONT COVER
This procedure covers the removal of both the right and left trim covers.
1. Turn power off by turning circuit breaker OFF.
2. Remove front cover from unit by pulling corners of cover from ball stud retainer
sockets on mainframe. See Illustration 5-2.
3. Remove rear cover by removing two screws retaining it on base assembly.
4. Disconnect hand switch on right side trim cover.
5. Remove two 4-40 x 3/8 inch binding head screws and one 6-32 x 1/2 inch black
oxide binding head screw holding each cover to base and mainframe assemblies.
6. Pull covers loose from five ball stud retaining sockets and carefully remove from
unit.
HAND SWITCH
BALL STUDS
5-2
(MODEL 2169360, 2236420 & 2275938 SERIES)
4-40 x 3/8
SCREWS
BATTERY COMPARTMENT COVER
SCREWS TOP AND BOTTOM
AMX-4+ SERVICE
DIRECTION 2173225-100
4-40 x 3/8
SCREWS
LEFT COVER
6-32 X 1/2 BLACK
OXIDE SCREW