Disconnect And Remove Batteries - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 6-2
SENSE LEADS ON BATTERY TEST TERMINAL STRIP (46-321370G1)
PRINTED WIRING BD ASM
46-321370G1
CABLE #1
SENSE
LEAD
(J6)
(J5)
CABLE #8
SENSE LEAD
CABLE #3
SENSE LEAD
6 3

Disconnect and Remove Batteries

CABLE #7
SENSE LEAD
(J2)
CABLE #7
SENSE LEAD
(J4)
(J7)
(J8)
CABLE #1
SENSE LEAD
CABLE #2
SENSE LEAD
1. Disconnect Cable #2 from Battery #9 (-) and slide boot over cable terminal.
See Illustration 6-3.
2. Disconnect Cable #5 from Battery #9 (+), and slide boot over cable terminal.
3. Remove Battery #9, Insulator #8, and Insulator #1 in front of Batteries #7 and
#8.
4. Disconnect sense leads from J7 and J8 on left side of battery test terminal strip.
5. Disconnect Cable #3 from Battery #7 (+) and slide boot over cable terminal.
6. Disconnect Cable #5 from Battery #8 (-) and slide boot over cable terminal.
7. Disconnect one of the Cable #1s from Battery #7 (-) and Battery #8 (+), and
slide a boot over each cable terminal.
8. Remove Batteries #7 and #8 and the separator pad between them.
9. Remove the remaining Insulator #1.
6-3
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
CABLE #4
SENSE LEAD
CABLE #6
SENSE LEAD
(J1)
BATTERY SENSE
HARNESS
(IF PRESENT)
CONNECTS TO J11
(J3)
CABLE #5
SENSE LEAD
AMX-4+ SERVICE
(J11)
(J10)
(J9)