GE MEDICAL SYSTEMS
REV 5
10 2
Cord/Cord Reel Installation
10 2 1 Cord Reel Installation (Domestic Models)
10 2 2 Cord Reel Installation (IEC Models)
1. Shut power off by turning circuit breaker OFF.
2. Feed cord reel leads through grommet in low voltage power supply housing. See
Illustration 10-1.
3. Install cord reel on side of low voltage power supply housing and secure with two
hex washer head screws and lock nuts.
4. Install cable grommet in cut-out in mainframe with two cap screws.
5. Connect three leads from cord reel to F1 (Black), F2 (White) and ground stud
(Green).
6. Slide low voltage power supply into mainframe housing and fasten with four hex
nuts.
7. Close top cover. Refer to Section 5-6.
1. Shut power off by turning circuit breaker OFF.
2. Feed cord reel leads through strain relief bracket and tighten bracket retaining
screw. See Illustration 10-2.
3. Install cord reel on side of low voltage power supply housing and secure with two
hex washer head screws and lock nuts.
4. Install cable grommet in cut-out in mainframe with two cap screws.
5. Connect three leads from cord reel to main terminal block.
- Use color code as identified by tag adjacent to terminal block.
6. Close top cover. Refer to Section 5-6.
10-3
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE