GE MEDICAL SYSTEMS
REV 5
7 5
Remove Motor and Brake
Note:
ILLUSTRATION 7-8
MOTOR AND BRAKE WIRING
MOTOR
BRAKE
6. Slide the support angle with its wheel block onto the wheel axle as far as it will
go.
7. Tighten the four locknuts and flat washers holding the wheel support angle to the
base assembly. Torque to 50-70 pound-inches (5.6 to 7.8 N-m).
8. Secure the wheel and caliper assembly to the wheel mounting support. Torque
to 40-60 pound-inches (4.5 to 6.7 N-m).
9. Install splash guard and secure with two lock nuts.
10. Install the motor and brake assembly. Refer to Section 6-10.
11. Lower the unit to the floor.
This procedure covers the removal of either the right or left motor and brake
assembly.
1. Move unit to a flat, level surface.
2. Shut off power.
3. Remove cassette drawer from unit. Refer to Section 5-1.
4. Disconnect motor and brake wire connectors from motor control panel. See Il
lustration 7-8.
CONNECTOR WIRES TO MOTOR CONTROL PANEL
7-8
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
TY–RAPS
AMX-4+ SERVICE