Install Latch Assembly and Components (Hall Effect Switch)
CAUTION
1. Shut off power by turning circuit breaker OFF.
2. Install solenoid in recess in latch block. See Illustration 5-4.
3. Install hex nut on top of solenoid.
Use care to avoid electrostatic discharge damage to the Hall effect sensor cir
cuit board.
4. Install Hall effect sensor circuit board and cover with two flat head screws.
5. Attach latch spring to latch with binding head screw.
6. Install latch in latch block.
7. Insert spring pin in latch block to retain latch in position.
8. Install latch block on spacer and fasten with four hex socket head capscrews.
9. Install latch assembly on mainframe and fasten with four hex socket head
capscrews.
10. Connect leads from latch assembly:
D
Hall Effect Sensor Board: three leads to terminal strip #1, #2 and #3.
D
Solenoid: two leads to terminal strip in accordance with labels on terminals.
11. If you moved low voltage power supply out of the way in Section 5 7, slide power
supply onto mainframe housing and fasten with four hex nuts.
12. Close top cover. Refer to Section 5-6.