Ap Assembly Installation - GE Healthcare Brivo XR118 Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Section 11.0

AP Assembly Installation

11.1
AP Assembly location selection
11.1.1
AP location selection process
1.) Place the AP at Default location and then perform wireless signal strength test. If passed,
2.) Place AP in Adjusting range and then perform wireless signal quality test. If passed, install AP
Page 122
6
EVISION
install AP at default location. Otherwise, continue Step 2.
Default Location
* Default Location 1-1: Only table bucky is required.
* Default Location 1-2: If wallstand is required.
at selected location. If failed to find a good location within Adjusting range, select any possible
location where wireless signal is good. Place AP at the location, then perform wireless quality
test. Compare wireless quality test results of all passed locations, select the best one to install
AP.
Section 11.0 - AP Assembly Installation
Adjusting Range
Default Location 3
Table
Default Location 1-1
Default Location 1-2
(Recommended)
(Recommended)
Figure 5-21 Room Layout
B
XR118 S
M
RIVO
ERVICE
Detector Insert Direction
< 4 meters
Exam Room