Horizontal Arm Brake and
Cable Assembly Removal
ILLUSTRATION 3-6
HORIZONTAL ARM OUTER,
CENTER AND INNER TUBES
INNER TUBE
CENTER TUBE
OUTER TUBE
LATCH BLOCK
CAPSCREWS
MOUNTING PLATE
CAPSCREW
6-32 X 3/8 HEX SOCKET
FLAT HEAD
1. Remove horizontal arm assembly. Refer to Section 3-5.
2. Position horizontal arm assembly with latch block up.
3. Extend Horizontal Arm until clearance hole on underside is exposed.
4. Remove cable clamp. See Illustration 3-6.
5. Disconnect Brake Cable connection.
6. Remove the two stop block capscrews securing the brake and cable assembly to
the end of the outer tube. See Illustration 3-6.
CABLE CLAMP
BRAKE CONNECTOR BRACKET
& GROUND WIRE
3-9
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
BRAKE & CABLE ASSEMBLY
BRACKET CAPSCREWS
6-32 X 1/2 HEX SOCKET HEAD
BRAKE CAPSCREWS
6-32 X .62 HEX SOCKET HEAD
AMX-4+ SERVICE
INNER SHAFT &
RETAINING RINGS
STOP BLOCK
CAPSCREWS
10-32 X 1/2 HEX
SOCKET FLATHEAD