Problem solving

M300 messages

The following table lists various messages that may display on the M300.
Priority
Message
None
All alarms paused
None
Audio paused
None
Default patient settings re-
stored
None
Discharged
None
High temperature reset
None
HR alarms off
None
HR,ASY,VF off
!
Low battery
238
Description
The alarm function of the
M300 device is temporarily
disabled. The message is
followed by minutes and
seconds which represent a
count-down timer.
The audible alarm signal for
the M300 device is silenced
for 2 minutes. The message
clears after the alarm si-
lence period times out.The
message is followed by
minutes and seconds which
represent a count-down
timer.
User settings were restored
to factory defaults.
The patient is discharged
(see page 110).
Reset due to the internal
temperature exceeding the
acceptable limit.
Alarms related to the heart
rate parameter are dis-
abled.
1
– Arrhythmia monitoring
is disabled.
– HR alarms are disabled
– The ASY/VF alarms fea-
ture is set to Follow HR
(not supported by the
M300).
The M300 battery charge is
low, with approximately
30 minutes of runtime re-
maining.
Instructions for use – Infinity
Action
Informational message,
no action required.
Informational message,
no action required.
Informational message,
no action required.
Informational message,
no action required.
Informational message,
no action required.
Informational message,
no action required.
Informational message,
no action required.
Recharge battery.
®
M300 and M300+ series – VG3.0