St Reference; Saving St Reference; Accessing The St Setup Page - Dräger Infinity M300 Series Instructions For Use Manual

Hide thumbs Also See for Infinity M300 Series:
Table of Contents

Advertisement

ECG, arrhythmia, and ST segment monitoring with M300

ST reference

You can save ST reference complexes as a reference point for future ST deviation measurement
comparisons. The first time you relearn ECG complexes, the current ST data are saved as a reference
data. This original ST reference data is updated each time you save ST references.
ST
Saving
reference
You can save the ST reference from the general ST complexes page and the single ST complex page
(see "ST complex pages" on page 163). Saving a reference complex in either page, saves all currently
displayed ST complexes as reference complex.
ST
To save
reference complexes
1 Open the BedView page by selecting the parameter box of the telemetry patient in the ICS Main
screen.
2 Select the ECG views button.
3 Select the ST complexes tab.
4 Select the desired ST complex panel.
5 Select the Save reference button or select X to close the ST complex panel.

Accessing the ST setup page

The ST setup page for a telemetry patient can only be accessed through the BedView page of the ICS.
1 Open the BedView page by selecting the parameter box of the telemetry patient in the ICS Main
screen.
2 Select the Telemetry/ Parameters button.
3 Select the Layout/ Parameters tab (if not already selected).
4 Select the ST tab.
168
Instructions for use – Infinity
®
M300 and M300+ series – VG3.0

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity m300+ series

Table of Contents