St Measuring And Iso Points; Adjusting St Measuring And Iso Points - Dräger Infinity M300 Series Instructions For Use Manual

Hide thumbs Also See for Infinity M300 Series:
Table of Contents

Advertisement

A ST complexes tab
B ST label (unique for each ST lead)
C Scale
D Waveform area
E Save reference button
F Reference button (on, off)
G Undo button
H Size [mV] selection control
I
Set points button
J ST measuring point selection controls
K ISO point adjustment controls

ST measuring and ISO points

You can change the ST measurement point and the ISO point from the single ST complex setup page.
The setup buttons and controls for changing the measuring points are located on the bottom of the page.
Changing the measuring point for one complex adjusts the measuring points for all ST complexes.

Adjusting ST measuring and ISO points

Whenever you adjust the ISO point and ST measurement point, the ST deviation is recomputed.
ST
To change
measurement points
1 Open the BedView page by selecting the parameter box of the telemetry patient in the ICS Main
screen.
2 Select the ECG views button.
3 Select the ST complexes tab.
4 Select the desired ST complex panel.
5 Select the arrows next to the ISO value to move cursor and adjust ISO point.
6 Select the arrows next to the ST value to move cursor and adjust ST point.
7 Select the Set points button to accept the settings and send them to the M300. Select X to close the
ST complex panel without saving changes.
Instructions for use – Infinity
ECG, arrhythmia, and ST segment monitoring with M300
®
M300 and M300+ series – VG3.0
167

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity m300+ series

Table of Contents