Dräger Infinity M300 Series Instructions For Use Manual page 273

Hide thumbs Also See for Infinity M300 Series:
Table of Contents

Advertisement

Heart rate level alarm adjust-
ment range
Monitoring frequency response
Sweep speed
Detected events/rhythms
Time to alarm for the two ven-
tricular tachycardia waveforms
per 60601-2-27 with amplitude
normal, one half and twice the
indicated value are:
Degree of protection against
electrical shock
Defibrillation protection
Electrosurgery and cautery
Instructions for use – Infinity
Infinity M300
20 to 300 bpm
0.5 to 40 Hz: ±3 dB
25 mm/s ±10 %
Asystole, Ventricular Fibrillation, Ventricular Tachycardia, Brady-
cardia, Ventricular Run, Accelerated Idioventricular Rhythm, Su-
praventricular Tachycardia, Ventricular Couplet, Ventricular Bi-
geminy, Tachycardia, Pause, Artifact, PVC/min
B1 aami4a
B1 aami4a_h
B1 aami4a_d
B1 aami4b
B1 aami4B_h
B1 aami4b_d
Type CF
In accordance per IEC 60601-2-27
Not intended for use during ESU procedures
®
M300 and M300+ series – VG3.0
Technical specifications
3 - 3.5 seconds
2 - 2.5 seconds
4 - 5 seconds
2 - 2.5 seconds
2.5 - 3 seconds
2 - 2.5 seconds
273