M300/M300+ network error message

The following error message may display on the M300/M300+:
Priority
Message
None
Network config error

PSK security credential expiration

The PSK security credential is valid for 2.5 years by default. The M300/M300+ will generate an expiration
warning when the device is in Secure mode, and the security credential is about to expire.
Warning
90-day warning
7-day warning
1-day warning
If you cannot admit an M300/M300+ patient to the ICS, refer to the clinical log for additional information
and messages.
Instructions for use – Infinity
Alarm
A 90-day warning will be issued when the security
credential is within 90 days of its expiration. The
user will receive a single-shot advisory tone, and a
persistent low-priority alarm banner in the header
bar of the screen.
A 7-day warning will be issued when the security
credential is within 7 days of its expiration. The user
will receive a single-shot advisory tone, and a
persistent medium-priority alarm banner in the
header bar of the screen.
The user will not be allowed to admit a patient on the
M300/M300+.
If the M300/M300+ travels out of wireless range
while the patient is still admitted during the 7-day
warning period, the device will re-connect when
returned to the wireless range.
A 1-day warning will be issued when the security
credential is within 1 day of its expiration. The user
will receive a single-shot advisory tone, and a
persistent high-priority alarm banner in the header
bar of the screen.
The user will not be allowed to admit a patient on the
M300/M300+.
®
M300 and M300+ series – VG3.0
Description
Multicast traffic is detected
on the VLAN.
Cybersecurity
Action
Contact your hospital's
biomedical and IT
departments for support.
Action
Contact your hospital's
biomedical department,
DrägerService, or
specialized service
personnel.
Contact your hospital's
biomedical department,
DrägerService, or
specialized service
personnel.
Contact your hospital's
biomedical department,
DrägerService, or
specialized service
personnel.