Dräger Infinity M300 Series Instructions For Use Manual page 198

Hide thumbs Also See for Infinity M300 Series:
Table of Contents

Advertisement

Recordings/reports
Digit
Description
5
Patient category/
QRS classification
6
Leads available
for processing
7
VTach count
8
VTach rate
9
SVT count
10
SVT rate
11
TACH count
12
TACH rate
13
BRADY rate
14
PAUS rate
15
HR source
16
RESP mode
198
Possible
Definition
values
<Space>
Adult, neither lead completed learning
1
Adult, ECG1 lead completed learning
2
Adult, ECG2 lead completed learning
B
Adult, ECG1&2 lead completed learning
n
Neonate
No valid lead to process
0
1
ECG1 is valid to process
2
ECG2 is valid to process
3
ECG1 & ECG2 are valid to process
5-F
Value = VTach count
(where A-F corresponds to 10-15)
0-A
Value = (VTach rate - 100)/10
(where A corresponds to 10)
3-A
Value = SVT count
(where A corresponds to 10)
0-A
Value = (SVT rate - 100)/10
(where A corresponds to 10)
5-F
Value = TACH count
(where A-F corresponds to 10-15)
0-A
Value = (TACH rate - 100)/10
(where A corresponds to 10)
0-F
Value = (BRADY rate - 30)/5
(where A-F corresponds to 10-15)
0-5
Value = (PAUS rate - 1.0)/0.5
(where A-F corresponds to 10-15)
E
ECG is heart rate source
P
ART is heart rate source
S
SpO2 is heart rate source
O
RESP monitoring off
M
Manual
A
Automatic
<Space>
N/A