Configuring arrhythmia alarm settings

In the following steps, the letters in parentheses refer to the diagram of the Alarms > Arrhythmia page
(see "Alarm setup for arrhythmia parameters" on page 130).
For alarm ranges and defaults for the M300, see "M300 alarm ranges and defaults" on page 134.
To configure the arrhythmia alarm settings of a bedside monitor or M300
1 Select the Alarms button in the menu bar of the BedView page.
2 Select the Arrhythmia tab (if not already selected).
3 Select the desired arrhythmia mode using the ARR mode buttons (A).
4 Select the corresponding button in the Alarm column (C) to select the alarm priority (High, Medium,
or Low) or to disable alarm monitoring. A crossed-out triangle
alarms settings page when alarm monitoring is disabled. The priority for asystole (ASY) and ventricular
fibrillation (VF) events cannot be changed. The alarm priority High is always assigned to these cate-
gories.
5 Select the corresponding setup button in the Rate column (D) to set the rate.
6 Select the corresponding setup button in the Count column (E) to set the count.
7 Use one of the following settings in the Archive column (F) to determine what happens in response to
an alarm:
– Off – no event is stored and no recording is generated.
– Store – stores the event in the disclosure database.
– Record – generates a strip recording.
– Store/Rec. – generates a strip recording and stores the event in the disclosure database.
8 Repeat steps 4 through 7 for each parameter you want to configure.
9 Select the Apply button.
Instructions for use – Infinity
®
M300 and M300+ series – VG3.0