Alarms

M300 alarm ranges and defaults

Parameter ranges and defaults

Parameter
Alarm limit range
HR
Upper: 25 to 300 bpm
Increment:
5 bpm
Lower: 20 to 295 bpm
PLS
Upper: 35 to 240 bpm
Increment of
Lower: 30 to 235 bpm
5 bpm
PVC/min
Upper: 1 to 60
Lower: not applicable
Increment of 1
SpO2
Upper: 21 to 100 %
Lower: 20 to 99 %
Increment: 1 %
ST
Upper: –14.9 to +15.0 mm
Increment:
0.1 mm or
Lower: –15.0 to +14.9 mm
0.01 mV
STVM/STCVM
Upper: 0.1 to 45.0 mm
Increment:
0.1 mm or
Lower: 0.0 to 44.9 mm
0.01 mV
134
Upper limit

defaults

120 bpm
(adult)
150 bpm
(pediatric)
120 bpm
(adult)
150 bpm
(pediatric)
10 (adult, pe-
diatric)
100 % (adult,
pediatric)
+2.0 mm
–1.49 to +1.50 mV
(+0.2 mV)
–1.50 to +1.49 mV
+2.0 mm
0.01 to 4.50 mV
(+0.20 mV)
0.00 to 4.49 mV
Instructions for use – Infinity
Lower limit
Alarm
defaults
default
45 bpm
On
(adult)
50 bpm
(pediatric)
45 bpm
Off (adult)
(adult)
On (pediatric)
50 bpm
(pediatric)
Not applicable Off
90 % (adult,
Off (adult)
pediatric)
On (pediatric)
–2.0 mm
Off
(–0.2 mV)
0.0 mm
Off
(0.0 mV)
®
M300 and M300+ series – VG3.0
Alarm
archive
default
Store/Rec.
Off
Off
Off
Off
Off