9.5.2 Control with 1 electrode and 1 switch
Electrode signal to the OPENING electrode.
Opening:
At constant speed, as long as the switch is operated.
Closing:
Control
mode 5
Digital
Sustained electrode signal
Control
Adjustment regulator B
Display of the MyoSelect (757T13)
9.5.3 Control with 1 switch
This control mode can be used with any MyoBock switch.
At constant speed, as long as the OPEN-side of the switch is operated. The
Opening:
MyoHand VariPlus Speed remains open.
At constant speed, as long as the CLOSE-side of the switch is operated.
Closing:
Control
mode 5
Digital
MyoHand VariPlus Speed
Control
opens as long as the OPEN-
side of the switch is oper-
ated.
Adjustment regulator B
Display of the MyoSelect (757T13)
32 | Ottobock
OPEN
Adjustment regulator B is used to adjust the speed of the "OPEN-
ING / CLOSING" function and the speed of grip force build-up.
Notice: Always check the adjustments on the prosthesis and
adapt them to the amputee.
OPEN
Adjustment regulator B is used to adjust the constant speed
of the "OPENING / CLOSING" function and the speed of grip
force build-up.
Notice: Always check the adjustments on the prosthesis and
adapt them to the amputee.
CLOSE
Signal from the switch
Grip force:
Duration of the signal
Adjustment regulator B
CLOSE
MyoHand VariPlus Speed
closes as long as the
CLOSE-side of the switch is
operated.
Grip force:
Duration of the signal
Adjustment regulator B
Indication
For amputees
with only one
muscle and
weak elec-
trode signal
Indication
For amputees
with weak or
no electrode
signals
8E38=9, 8E39=9, 8E41=9