Before Performing A Dka Calibration - Thermo Scientific Orion 2120XP User Manual

Calcium hardness analyzer
Table of Contents

Advertisement

Calibration
Before Performing
a DKA Calibration
Recommendations
Pipet Dispensing
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
Maximum system accuracy is ensured through a fast, easy and accurate
calibration performed in the expected sample range. This calibration
procedure uses equipment supplied in the calibration kit
(Cat. No. 2120XPCAL).
Wait at least one hour after changing the reagent or cleaning electrode
before performing a calibration.
Check that the present sample concentration is equal or less than
standard 1. If not, a calibration can still be performed with a loss in
accuracy as long as background concentration is not significantly larger
than the concentration of standard 1. The best results are achieved
when standard 1 is twice the level of the present sample concentration.
Proper pipet technique is crucial to the calibration. If the operator is
not familiar with pipet equipment or technique, refer to the Appendix,
Pipet Operation section.
Do not touch the pipet tip with bare fingers. Risk of contamination is
possible. Use powder free gloves to prevent contamination.
When using a pipet to inject standards during calibration, use a fresh
pipet tip for each standard and each injection.
When a fresh tip is installed, rinse the tip with the same concentration
standard as the intended injection.
Fill the tip and dispense the standard 1 to 3 times before injecting the
standard into the flow cell. Be sure to discard the standard in the
drain between the rinses.
Wipe any excess liquid from the outside of the tip with out touching
the opening.
Insert the pipet tip into calibration port on flow cell cap.
Be sure to hold the pipet completely vertical while dispensing. Do not
inject the standard on the side of the flow cell, as a film may be retained
on the inside of the wall, resulting in errors in precision of calibration.
Figure IV-4
Refer to Figure IV-4.
IV-5