Maximum Heating
Rate Setting (LIMt)
Thermo Scientific
The higher the heating rate, the faster the heat up time but the wider the
temperature overshoot and undershoot. Therefore, if you want to reduce
the temperature fluctuation, you need to limit the heating rate to a certain
degree by limiting maximum heating rate. [Same function as Temperature
Setting].
If, for example, the current heating rate is 100% but you want to set the
heating rate limit at 60%, do as follows:
Step 1 : Press the CONFIG Button.
Step 2 : Select Limit (LIMt) turning and pressing the Stirrer Knob.
Step 3 : Change the limit to the desired value by turning the Stirrer Knob
and press the Stirrer Knob.
Step 4 : Save the desired limit by using the Stirrer Knob as shown below.
[When you need to save the changes]
If you don't want to save it, press
the Stirrer Knob when display
shows "SAVE NO". [When you
don't need to save the changes]