Changing Low Temp
Limits (t-L) cont.
Check Configuration
Settings (LISt)
Thermo Scientific
Step 4: Save the desired limit by using the Stirrer Knob as shown below.
[When you need to save the changes]
If you don't want to save it, press the Stirrer Knob when display shows
"SAVE NO". [When you don't need to save the changes]
Step 5 : Terminate the configuration mode as described in Configuration
Mode.
• The allowed range of the temperature limit setting is 0°C ~ 350°C.
• However, low temperature limit [t-L] set point should not be more than high
temperature limit [t-H] set point.
You can check the current configuration settings as follows:
Step 1: Press the CONFIG Button.
Step 2: Select the configuration checking mode [LiSt]
shown below by turning the Stirrer Knob.
Step 3 : Press the Stirrer Knob and check the current
settings by turning the Stirrer Knob.
Section 4
Control Panel
Digital Hotplate
4-22