Section 4
Control Panel
Auto-tuning (PID
Parameter Calibration
4-18
Digital Hotplate
Step 2: Select the offsetting mode(OFST or OFSP) by turning the Stirrer
Knob and press the Stirrer Knob as shown below.
[OFST- Top plate offsetting]
Step 3: Select the offset value by turning the Stirrer Knob.
The allowed ranges of the temperature offsetting are from -10 to +50°C for
OFST and from -3 to +3°C for OFSP.
[OFST- Top plate offsetting]
Step 4: Save the desired offsetting value by pressing the Stirrer Knob as
shown below. [When you need to save the changes]
[OFST- Top plate offsetting]
If you don't want to save it, press the Stirrer Knob when display shows
"SAVE NO". [When you don't need to save the changes]
[OFST- Top plate offsetting]
Step 5: Terminate configuration mode as described in Configuration Mode.
[OFSP- External temperature probe]
[OFSP- External temperature probe]
[OFSP- External temperature probe]
[OFSP- External temperature probe]