Temperature Offsetting (Ofst/Ofsp*) - Thermo Scientific HPS RT2 Advanced Manual

Table of Contents

Advertisement

Selection of the Timer
Mode (continued)
Temperature Offsetting
(OFST/OFSP*)
Thermo Scientific
If you don't want to save it, press the Stirrer Knob when display shows
"SAVE NO". [When you don't need to save the changes]
Step 5: Terminate the configuration mode as described in Configuration
Mode.
• Refer to Top plate's temperature influence on Heating bath and Erlenmeyer
flask's temperature for understanding of temperature relation between top
plate and samples.
The temperature shown on the Actual Temperature Display is measured by
a temperature sensor inside the unit. However, this temperature can be
different from the temperature of the external temperature probe. If
needed, you can offset such temperature differences at 0.1°C interval.
* Offsetting of top plate is displayed as OFST and offsetting of the exter-
nal temperature probe is displayed as OFSP.
• The allowed range of the temperature offsetting:
OFST (-10 ~ +50°C) / OFSP (-3 ~ +3°C)
• See the graph in Accessories which shows the temperature difference
depending on the top plate temperature.
• Refer to Temperature Offsetting for the relation between top plate's actual
temperature and the shown temperature on the display in case of the unload
condition.
If, for example, the actual temperature of the top plate is 100°C but the
displayed temperature is 95°C, you can match the displayed temperature
with the actual temperature of the unit by selecting the offset value of
+5°C and save it as described below:
Step 1 : Press the CONFIG Button.
Section 4
Control Panel
Digital Hotplate
4-17