Maximum Heating Rate
Thermo Scientific
The higher the heating rate, the faster the heat up time but the wider the
Setting
temperature overshoot and undershoot. Therefore, if you want to reduce
the temperature fluctuation, you need to limit the heating rate to a certain
degree by limiting maximum heating rate. [Same function as 4.3.4]
If, for example, the current heating rate is 100% but you want to set the
heating rate limit at 50%, do as follows:
Step 1: Press the CONFIG Button. The default value is 100%.
Step 2 : Change the limit to the desired value by turning the Stirrer Knob
and press the Stirrer Knob.
Step 3 : Save the desired limit by using the Stirrer Knob as shown below.
[When you need to save the changes]
If you don't want to save it, press the Stirrer Knob when display shows
"SAVE NO". [When you don't need to save the changes]
Step 3: Terminate the configuration mode as described in Configuration
Mode.