If you see below "Error codes", turn off the power. Restart after having the
unit rest enough.
Error code
Cause
ERRO NO 1
Related to Heater system
ERRO NO 2
ERRO NO 3
ERRO NO 4
ERRO NO 5
Related to temperature
ERRO NO 6
probe
ERRO NO 7
if Top plate is overheated
if sensor detects a sudden temperature
changes
if PCB is overheated
if sensor is not connected correctly
if temperature difference has arisen
between two sensors
if temperature probe is in the air
temperature probe is not soaked enough
into media
Hotplate & Magenetic Stirrer
Section 7
Maintenance
Effect
Heater off
Heater off
Heater off
Heater off
Heater off
Heater off
Heater off
7-3