Coefficient (COEF)
continued
Thermo Scientific
[User-selectable temperature control modes]
Provide the fastest heat up time but the widest fluctuation range of
Fast Mode
temperature overshoot and undershoot
Provide the optimal balance between the heat up time to reach the
Optimal Mode
target temperature and the allowed fluctuation range of temperature
overshoot and undershoot (factory default)
Provide the slowest heat up time but the narrowest fluctuation range
Slow Mode
of temperature overshoot and undershoot
User Mode
Allow auto-tuned parameters to be used for temperature control
Allow linear heating up to the target temperature using the maximum
heating rate and switching on and off the heater based on the target
Point Mode
temperature (Note that this mode shows the widest fluctuation range
of temperature overshoot and undershoot.)
Step 4: Save the desired mode by
using the Stirrer Knob as shown
at right. [When you need to save
the changes]
If you don't want to save it, press
the Stirrer Knob when display
shows "SAVE NO". [When you
don't need to save the changes]
Step 5: Terminate configuration mode as describe in Configuration Mode.