Max. Heating Rate
Setting (LIMt) cont.
Changing the High
Temp Limits (t-H)
Thermo Scientific
Step 5: Terminate the configuration mode as described in Configuration
Mode.
• When the heat up time is too slow, check whether the maximum heating rate
is set too low.
• Maximum heating rate set point can only be applied in top plate's control
mode, not the external temperature probe mode.
Note that the default setting of the high temperature limit [t-h] 350°C. If
needed, however, you can set your own temperature limits. If, for example,
you want to set the high temperature limit at 200°C, do as follows:
Step 1 : Press the CONFIG Button.
Step 2 : select the temperature limit mode shown
below by turning the Stirrer Knob. Press the Stirrer Knob and check the
current high temperature limit:
Step 3 : Change the limit to the desired value by turning the Stirrer Knob.
Step 4 : Save the desired limit by using the Stirrer Knob as shown below.
[When you need to save the changes]
Section 4
Control Panel
Digital Hotplate
4-20

Advertisement

Table of Contents
loading