RANGE OF RADIOGRAPHIC PARAMETERS

PARAMETER
PARAMETER
kVp
From 40 kVp 150 kVp in 1 kV steps.
From 10 mA to 800 mA through the following mA stations:
mA
10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800.
(Depending on the Generator model)
mAs
Product of mA x Time values from 0.1 mAs to 500 mAs
From 1 to 10000 milliseconds through the following Time stations:
ms
1, 2, 3, 4, 5, 6, 8, 10, 12, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800,
1000, 1250, 1600, 2000, 2500, 3200, 4000, 5000. 6400, 8000 and 10000.
mAs: 0.1 mAs to 500 mAs
AEC
AEC
Exposure Time: Nominal shortest irradiation Time = 14 ms
Max ms
550, 1000.
with DR Detector
7.3.3

DUTY CYCLE

0171F EN 20181009
The Generator duty cycle is continuous, but limits should be set during
installation depending on the capacity of the X-ray Tube.
RANGE
RANGE
DX-D 300 U-Arm
User Manual
87