86 | DX-D 300 | Operating the DX-D Software Console

Radiographic Parameters

You can set up following radiographic parameters:
Figure 23: Radiographic parameters
To increase the radiographic parameters step by step, use the UP and DOWN
arrows. The values increase or decrease step by step each time the
corresponding button is touched.
• kVp: shows the radiographic kVp value (X-ray tube voltage) selected for
the parameters.
• mAs can show:
• The radiographic mAs value selected for the parameter.
• The max mAs for an exposure using AEC. Based upon the max mAs
• When an exposure is made, it shows the actual mAs at the end of the
• mA: shows the radiographic mA value (current) selected for the
parameter. The focal spot selection is changed automatically if the mA
value exceeds the applicable range.
• ms can show:
• The time value (in milliseconds) selected for the parameter.
• When an exposure is made, it shows the actual time at the end of the
• Max ms shows: the maximum exposure time allowed with DR detector
operation (550 ms or 1000 ms). Based upon this, the generator must limit
its maximum exposure time. X-rays outside the integration time slot of the
DR detector are NOT allowed. This makes that with AEC, the exposure is
0172G EN 20230627 1651
and the mA setting, the generator must limit its maximum exposure
time. X-rays outside these limits are NOT allowed. The highest allowed
setting for max mAs depends on the mA.
exposure.
exposure.