4.1.6 Setting CP calculation value conditions

Set the conditions necessary for CP calculation in this mode.
Caution on handling
• The mode is displayed when CP calculation value is selected for the output type.
Setting range
• CO concentration
• Furnace temperature setting : "800°C to 1000°C" (In steps of 1°C)
Initial value
• CO concentration
• Furnace temperature setting : "900°C"
1) Move the cursor to <7. CP calculation
Value Condition Setting> using the
key and then press the
screen is switched.
(1) CO concentration setting
Select <Measured Value> or <Regular Value>
as CO concentration required for calculation.
2) Move the cursor to <1. CO Concentration
Setting> using the
the
key, and the screen is switched.
3) Select measured value or regular value
using the
• Select the measured value and press the
key to return to the previous screen.
• Select the regular value and press the
key, and temporary CO concentration
value is displayed.
Select a digit using the
change the value using the
Press the
ous screen.
Caution on setting
If CO analyzer is not provided, setting the
CO concentration value to <Measured
Value> results in a calculation error. Be sure
to enter the value properly.
: Regular value; "10.0 to 40.0% CO" (In steps of 0.1%)
: "Regulated to 20.0% CO"
key, and the
key and then press
key.
key, and
key.
key to return to the previ-
Setting complete
ENT
,
(Measured
value)
- 34 -
ENT
ESC
,
ENT
ESC
,
(Regular
ENT
ESC
,
value)
INZ-TN2ZSVS-E