4.3.2 Average output time

Average corrected time can be set in this mode.
Caution on setting
• The message shown at right is displayed if average cor-
rected value is not selected in "4.3.1 Selecting output
type."
• Sampling cycle is 30 seconds.
• The average of the values from the current time to before
the time of setting is output at intervals of 30 seconds.
Accurate average value can be obtained when the set time
elapses after the data is input.
Setting contents
• "01 to 59 minutes (in steps of 1 minute)" or "01 to 04 hours (in steps of 1 hour)"
Initial value
"01 hour"
(1) Move the cursor to <2. Average time> using the
key, and press the
The screen switches as shown at right.
(2) Move the cursor to the channel (CH) to be set using
the
key.
Press the
be set.
(3) Move the cursor to "Hour" and "Minute" using the
key, and enter numeric values using the
Move the cursor to the "numeric value" using the
key again, change the value using the
then press the
(4) Make the setting for other channels in the same man-
ner.
The setting has now been completed.
INZ-TN2ZSVS-E
key.
key, and the cursor moves to the time to
key.
key.
key, and
- 47 -
ENT
ESC
,
ENT
ESC
,
ENT
,
,
Setting complete