Fuji Electric ZSVS-2 Instruction Manual page 36

Compact type gas analyzer
Table of Contents

Advertisement

Zero calibration of O
1. When <Dry N2> is selected in zero calibration mode
(1) Calibration is performed simultaneously with the infrared ray sensor.
See 4.1.1 (1) Zero calibration mode.
2. When <Air> is selected in zero calibration mode
CAUTION
• The O
zero calibration screen is displayed when
2
the O
analyzer is added.
2
(1) Press the
key to perform zero calibration.
CAUTION
• If "ON" is selected for O
tion of O
analyzer is not performed in this mode.
2
Follow the procedure shown below to perform zero
calibration of the O
(2) The screen is automatically
switched to the span calibration
screen of the 1st component (Ch1).
(3) Select "both" or "each," and then
press the
key to carry out span
calibration. See 4.1.1 (2) Span
calibration mode.
(4) The screen switches to "Next Ch span calibration" or "O
zero calibration."
The number of span calibration screen varies depending
on specifications.
(5) Feed the span gas that does not contain N
calibration gas inlet at the rate of 0.5L/min.
Example: Continuously feed the span gas fed immediately
before.
(6) When the display stabilizes in 2 to 3 minutes, press the
key.
(7) When calibration is completed, the O
screen appears.
(8) Perform O
span calibration in the same manner as other
2
span calibrations.
analyzer
2
analyzer, zero calibra-
2
analyzer.
2
Span calibration is carried
out repeatedly for multiple
component specifications
without O
or O
2
span calibration
2
- 24 -
MODE
analyzer.
2
In the case of O
2
2
from the
2
ENT
ESC
Auto
ENT
ESC
analyzer only
ENT
ESC
INZ-TN2ZSVS-E