Fuji Electric ZSVS-2 Instruction Manual page 96

Compact type gas analyzer
Table of Contents

Advertisement

Standard Requirements for Sample Gas
0.5 L/min ±0.2 L/min
• Flow rate:
• Temperature:
0 to 40˚C at inlet of sampling gas
10 to 70˚C at tip of non-fixed type probe
(available at option)
70 to 400˚C at tip of fixed type probe
(available at option)
• Pressure:
0 to 2 kPa (Gas shall be discharged into
atmospheric air.)
• Dust:
10 mg/Nm
• Mist:
Unallowable
• Corrosive gas: HCl 10 ppm or less
Others Unallowable
• Standard gas for calibration:
Zero gas; N
Span gas; Concentration limited within 90
Options
• Gas extractor: Used for aspirating sample gas.
Non-fixed type; Since this type is used
Fixed type; Used for continuous measure-
Material; SUS316
• Sample inlet tube:
Used for delivering gas from the extrac-
tor to sampling block.
Shape;
Material; Teflon
Installation Requirements
• Selection of a place which does not receive direct sunlight,
rain, wind nor radiation from hot substances.
If such a place cannot be found, a roof or cover should be
prepared for protection.
• Avoidance of a place under heavy vibration
• Selection of a place where atmospheric air is clean
• Discharge of exhaust gas into atmospheric air at a safe
location
• Avoidance of use in an explosion-proof area
3
or less
or clean air
2
However, clean air cannot be
used if CO
and O
are included
2
2
in sample gas components.
to 100% of the range of each
sample gas component.
Unusable at concentrations
beyond 100%.
for intermittent measure-
ment, it cannot be fixed.
Material;
SUS304/polypropylene
ment. Flange 5K25A FF
Sampling pipe length select-
able among 300, 400, 600
and 800mm
φ6/φ4 × 5 m or φ6/φ4 × 10 m
Scope of Delivery
• Gas analyzer system
• Standard accessories (Refer to the table at top Following
table.)
• Instruction manual
Item to be Prepared Separately
• Standard gas (ZBM) and pressure regulator (ZBD)
• Recorder (when necessary, Fuji's product type PHR)