This product is not explosion-proof. When handling dangerous gas, adequate attention shall be paid.
Standard Specifications
• Measuring system:
CO
, CO and CH
2
red absorption method with single light
source and single beam (single beam
method)
O
; Galvanic cell method
2
• Measurable component and min./max. measuring range:
CO
; 0 to 200 ppm / 0 to 100 %
2
CO; 0 to 200 ppm / 0 to 100 %
CH
; 0 to 1000 ppm / 0 to 100 %
4
O
; 0 to 5 % / 0 to 25 %
2
Max. 4 components measurable includ-
ing O
2
• Number of measuring ranges:
3 ranges
• Max. range ratio 1:5
• Warm-up time: 30 min after power-on
Provided with count-down timer indicat-
ing function.
• Analog output:
In up to 8 channels.
4 to 20 mA DC or 0 to 1 V DC (linear)
Non-isolated output
Allowable load; 4 to 20 mA DC, 550Ω or
• Instantaneous value output of each gas
component
• Instantaneous value output after O
correction (when provided with O
lyzer)
• Average value output after O
tion (when provided with O
• CP calculation value output (when pro-
vided with CO
* The channel numbers of indicated value
and output value correspond to each
other one by one.
* An exclusive 25-pin cable is standard-
equipped.
• Communication output:
RS-232C Modbus protocol
* Use a commercially available product
(D-sub 9-pin cable).

8. SPECIFICATION

8.1 Specification
; Non-dispersive infra-
4
less
0 to 1 V DC, 100 kΩ or
more
2
ana-
2
correc-
2
analyzer)
2
analyzer)
2
- 82 -
• Indicated values:
Digital 4-digit indication (by LCD with back
light)
• Instantaneous values of respective gas
components
• Instantaneous values after O
(when provided with O
• Average value after O
provided with O
• CP calculation value display (when pro-
vided with CO
* The channel numbers of indicated value
and output value correspond to each
other one by one.
• Power supply:
Rated voltage;
100 to 115 V AC or 200 to 240 V AC
Working voltage;
85 to 132 V AC or 180 to 264 V AC
* Depending on customer's code selection.
Rated frequency; 50/60 Hz
Max. rated power; 150 VA
Inlet; Class 1 type conforming with EN60320
• Operating conditions:
Ambient temperature; 0 to 40˚C
Ambient humidity; 90% RH or less
* Condensation unallowable
• Storage conditions:
Ambient temperature; -20 to 60˚C
Ambient humidity; 95% RH or less
* Condensation unallowable.
• External dimensions (H × W × D mm):
211 × 365 × 527
• Mass:
Approx. 12 kg
• Finish color:
Cover; White pearl mica
Base; Medium gray metallic
• Enclosure design:
Casing made of steel plates for indoor
installation.
• Gas-contacting part materials:
Gas inlet/outlet; Polypropylene
Sample cell; SUS304/neoprene rubber
Transparent window: CaF2
Internal pipes: Toalon tube/Teflon tube
• Gas inlet/outlet: φ6/ φ3 hose end
• Purge gas flow rate:
1 L/min (to be purged as required)