Output Hold - Fuji Electric ZSVS-2 Instruction Manual

Compact type gas analyzer
Table of Contents

Advertisement

4.2.3 Output hold

Output signals are put to hold during calibration in measurement (zero and span calibrations), and gas
change.
Setting contents
• ON
: Holds output signals.
• OFF : Resets the hold of output signals.
1. Calibration and hold operations in measurement
Measurement operation
Zero/span calibration
Output hold
(ON setting)
1. Change time after calibration
2. Zero/span calibration
Note 1:
Gas change : On completion of calibration in measurement, sample gas is fed and CHANGE is displayed.
2. Hold operation in standby mode
Output is not put to hold irrespective of hold "ON/OFF" setting.
Calibration
operation
Measurement
(zero, span)
start
MEAS
ON
OFF
ON
OFF
ON
OFF
: Keeps holding.
: Holds.
Initial value
• OFF
Change
time
OFF
OFF
- 40 -
The end of
measurement
MEAS
OFF
INZ-TN2ZSVS-E