4.2.3 Output hold
Output signals are put to hold during calibration in measurement (zero and span calibrations), and gas
change.
Setting contents
• ON
: Holds output signals.
• OFF : Resets the hold of output signals.
1. Calibration and hold operations in measurement
Measurement operation
Zero/span calibration
Output hold
(ON setting)
1. Change time after calibration
2. Zero/span calibration
Note 1:
Gas change : On completion of calibration in measurement, sample gas is fed and CHANGE is displayed.
2. Hold operation in standby mode
Output is not put to hold irrespective of hold "ON/OFF" setting.
Calibration
operation
Measurement
(zero, span)
start
MEAS
ON
OFF
ON
OFF
ON
OFF
: Keeps holding.
: Holds.
Initial value
• OFF
Change
time
OFF
OFF
- 40 -
The end of
measurement
MEAS
OFF