Resetting Average Output Value - Fuji Electric ZSVS-2 Instruction Manual

Compact type gas analyzer
Table of Contents

Advertisement

4.2.4 Resetting average output value

Integrated average output value and display can be reset as follows.
Caution on operation
• Immediately after the reset is "execute," the reading and the output value appear as 0ppm, vol%.
• Reset is allowed only for the instrument of specifications
with average value output provided. If average value
output is not provided, NO AVERAGE is displayed as
shown at right.
• See <Maintenance Mode> → <1. CH No.> and
<Maintenance Mode> → <2. Average Time> for the
setting of average value.
(1) Using the
Output>, and then press the
The cursor moves within the mode.
(2) Press the
select screen appears again.
Reset has now been completed.
Example of average operation and reset operation
When average value is set to 1 hour
• Sampling cycle is 30 seconds (fixed irrespective of the average time setting).
• The average of the values up to 1 hour before the current time is output at the intervals of 30
seconds.
• At the time of reset, all the values up to the current time are regarded as 0 for calculation. Conse-
quently, accurate average value can be obtained 1 hour after the reset.
key, move the cursor to <Reset Av.
key.
key, and reset is executed, and the item
Average value
Reset
1 hour
- 42 -
ENT
ESC
,
ENT
Setting complete
Time
INZ-TN2ZSVS-E