Fuji Electric ZSVS-2 Instruction Manual page 35

Compact type gas analyzer
Table of Contents

Advertisement

<In the case of Air>
Caution on operation
• If O
analyzer is provided, even if "ON" is
2
selected for O
zero calibration, calibration is
2
not performed because O
zero gas.
7) The cursor moves to
the reading.
Remove the pipe
connected to the
sample gas inlet.
MODE
CH1 span
Calibration screen
8) Press the
key, and the air is
ENT
automatically sucked from the
sample gas inlet. Check that the
flow rate reading of the flow
checker is kept
at approximately
0.5 L/min.
MODE
CH1 span
Calibration screen
9) After arbitrarily set time elapses,
the CH1 span calibration screen appears.
Caution on operation
• Zero calibration is performed in batch for all the ranges of selected components.
• If "OFF" is selected for zero calibration of all the components, all zero calibrations are not performed.
• Zero calibration is performed at the time arbitrarily set. The time of completion of zero calibration
is displayed at the upper right-hand corner of the screen.
How to forcibly perform zero calibration
To forcibly perform zero calibration without waiting for the arbitrarily set time of calibration to elapse,
press the
key in 8).
INZ-TN2ZSVS-E
is contained in the
2
ENT
ESC
,
,
ENT
ESC
MODE
- 23 -
<In the case of Dry N
2
7) The cursor moves to
the reading.
MODE
CH1 span
Calibration screen
8) Connect N
gas to the calibration
2
gas inlet. Press the
ENT
check that the flow rate reading
of the flow checker is kept at
approximately
0.5 L/min.
MODE
CH1 span
Calibration screen
Auto
>
ENT
,
,
key and
ENT
ESC
ESC