Fuji Electric ZSVS-2 Instruction Manual page 38

Compact type gas analyzer
Table of Contents

Advertisement

Operation in the case where "each" is selected
1) Select "each" for calibration operation using the
key, and press the
2) The cursor and concentration value are displayed
only for the target range.
Span calibration concentration value is also
displayed at the bottom of the screen.
3) Change the span calibration concentration value
using the
key.
4) Feed the span gas for target of calibration.
Adjust the flow rate at about 0.5L/min.
5) When the reading stabilizes in about 2 to 3
minutes, press the
6) The mode automatically switches to the Ch span
calibration mode.
7) To calibrate other ranges, press the
repeat the procedure in 3) to 6)
8) Follow the procedure in 3) to 7) to carry out span
calibration of all the components.
Caution on operation
• To change span gas calibration concentration
value, select <Menu Mode> → <2. Setting
about Calibration> → <2. Span Calibration
Value>.
Calibration has now been completed.
key.
key.
key, and
- 26 -
ENT
,
MODE
ENT
,
Calibration complete
ESC
ESC
INZ-TN2ZSVS-E